dc.identifier.citationNaughton, David, An in vitro study evaluating the efficacy of a novel mount with torque control designed to tighten Osstell SmartPegs, Trinity College Dublin, School of Dental Sciences, Dental Science, 2022en
dc.identifier.otherYen
dc.descriptionAPPROVEDen
dc.description.abstractObjectives: The aim of this study was to evaluate the effects that different transducer tightening methods have on implant stability quotient (ISQ) values. A secondary aim was to determine the reliability of ISQ measurements from implants of varying connections and geometries, in various bone densities with different transducer tightening methods. Materials and methods: Fifty six implants, comprising 7 different types of implant, were placed in 8 polyurethane blocks representing D1, D2, D3 and D4 bone densities. Resonance frequency analysis (RFA) transducers were attached to each implant in four different ways: hand tightening with gauze, hand tightening with a plastic wrench, hand tightening with a novel mount with inbuilt torque control and tightening to 6 Ncm with a calibrated torque wrench. ISQ measurements were performed using two Osstell devices, from a buccal and mesial direction and repeated by a second operator. An intra-class correlation coefficient (ICC) was calculated to assess the reliability between the two RFA devices, buccal and mesial directions and operator one and two. Linear mixed effects regression was used to determine the effect explanatory variables had on ISQ values. Results: There was a statistically significant difference in ISQ values obtained by hand tightening transducers with gauze compared to the calibrated torque wrench. There was excellent agreement between the two RFA devices (ICC 0.986) and between buccal and mesial measurements (ICC 0.977). For all transducer tightening methods there was excellent inter-operator agreement for D1 and D2 (ICC >0.8), good agreement for D3 (ICC >0.7) but very poor agreement for D4 (ICC 0.04-0.2). Bone density accounted for 36% of the variation in ISQ values, the implant accounted for 11% and the operator for 6%. Conclusions: Transducer tightening method has a statistically significant impact on ISQ values. However, the greatest impact on ISQ values is due to bone density. Furthermore, reliability of RFA is very poor in soft bone.en
dc.publisherTrinity College Dublin. School of Dental Sciences. Discipline of Dental Scienceen
dc.rightsYen
dc.subjectResonance frequency analysisen
dc.subjectDental implantsen
dc.titleAn in vitro study evaluating the efficacy of a novel mount with torque control designed to tighten Osstell SmartPegsen
dc.typeThesisen
dc.type.supercollectionthesis_dissertationsen
dc.type.supercollectionrefereed_publicationsen
dc.type.qualificationlevelDoctoralen
dc.type.qualificationnameProfessional Doctor of Dental Surgery (D.Ch.Dent)en
dc.identifier.peoplefinderurlhttps://tcdlocalportal.tcd.ie/pls/EnterApex/f?p=800:71:0::::P71_USERNAME:DANAUGHTen
dc.identifier.rssinternalid246911en
dc.rights.ecaccessrightsopenAccess
dc.identifier.urihttp://hdl.handle.net/2262/101370


Files in this item

Thumbnail
Thumbnail

This item appears in the following Collection(s)

Show simple item record