dc.identifier.citation | Nora May Nuala Crotty; Marie Boland; Nick Mahony; Bernard Donne; Neil Fleming, Reliability and Validity of the Lactate Pro 2 Analyser, Measurement in Physical Education and Exercise Science, 2021,25, 3, 202 - 211 | en |
dc.identifier.other | Y | |
dc.description.abstract | This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n = 258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L−1
for blood lactate (BLa) ≤4.0 mmol.L−1 to −1.6 to +1.4 mmol.L−1 for BLa >8.0 mmol.L−1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa ≤ 8 mmol.L−1, and using the LP2 to identify training intensities at lactate threshold. | en |
dc.format.extent | 202 | en |
dc.format.extent | 211 | en |
dc.language.iso | en | en |
dc.relation.ispartofseries | Measurement in Physical Education and Exercise Science; | |
dc.relation.ispartofseries | 25; | |
dc.relation.ispartofseries | 3; | |
dc.rights | Y | en |
dc.subject | YSI 1500 Sport | en |
dc.subject | Portable lactate analyzers | en |
dc.subject | Lactate threshold | en |
dc.subject | Bland- Altman limits of agreement | en |
dc.subject | Dmax | en |
dc.title | Reliability and Validity of the Lactate Pro 2 Analyser | en |
dc.type | Journal Article | en |
dc.type.supercollection | scholarly_publications | en |
dc.type.supercollection | refereed_publications | en |
dc.identifier.peoplefinderurl | http://people.tcd.ie/nflemin | |
dc.identifier.peoplefinderurl | http://people.tcd.ie/njmahony | |
dc.identifier.rssinternalid | 218863 | |
dc.rights.ecaccessrights | openAccess | |
dc.subject.TCDTag | Physiology | en |
dc.subject.TCDTag | Sports Science | en |
dc.identifier.rssuri | https://doi.org/10.1080/1091367X.2020.1865966 | |
dc.status.accessible | N | en |
dc.identifier.uri | http://hdl.handle.net/2262/98152 | |